Wika 632.51+8xx Manual do Utilizador Página 5

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 52
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 4
WIKA Operating instructions pressure gauges Model 632.51 with 831 per ATEX
5
11059738 11/2008 GB/D/F
GB
3. Technical data and intended use
3. Technical data and intended use
Working pressure
Steady: full scale value
Fluctuating: 0.9 x full scale value
Overpressure safety: 50 x scale range 7 bar maximum
Pressure connection
n According to the general technical regulations for pressure gauges,
respectively (i.e. EN 837-2 “Selection and installation recommendations for
pressure gauges”).
When screw-fitting the gauges the force required for sealing must not be ap-
plied through the case or terminal box but, using a suitable tool, only through
the spanner flats provided for this purpose at the square of the connector.
Installation using
a spanner
Temperature effect
When the temperature of the measuring system deviates
from the reference temperature (+20 °C):
max. ±0.6 %/10 K of full scale value
IP Ingress protection
IP 54 per EN 60 529 / IEC 60 529
Vista de página 4
1 2 3 4 5 6 7 8 9 10 ... 51 52

Comentários a estes Manuais

Sem comentários