Specications Models 891.44 and 892.44 (Ex Version)
Power supply UB
for Non-Ex version DC
12 V < U
B 30 V
for Ex version see the section ‚Ex protection‘!
Permissible residual ripple
% of span/10 V
0.1
Supply voltage eect % ss
10
Output signal 4 … 20 mA, 2-wire system
Permissible max. load R
A for Non-Ex versions, Model 891.44:
R
A (UB - 12 V) / 0.02 A with RA in Ω and UB in Volt
for Ex versions, Model 892.44:
R
A (UB - 14 V) / 0.02 A with RA in Ω and UB in Volt
Eect of load % of span
0.1
Output signal adjustment
Zero point, electrical A
djustment of zero point through brief bridging of terminals 5 and 6,
or using the „scale selection switch“ option, selectable via button
1)
Scale selection 4 scales selectable via BCD switch
Linearity % of span
1.0 (limit point calibration)
Permissible
ambient temperatures °C -40 … +80, -40 … +60 with oxygen
Compensated temp. range °C -40 … +80
Temperature coecients in
compensated temp. range
Mean TK of zero
% of span/10 K
0.3
Mean TK of span
% of span/10 K
0.3
Ex protection according to EC-Type Examination Certicate
BSV 08 ATEX E 018 X for Model 892.44
Ex certication EEx II 2G EEx ia IIC T6
Conformity specications
Power supply U
i DC 14 … 30 V
Short circuit rating I
i mA max. 100
Rating P
i W max. 1
Internal capacitance C
i nF 12
Internal inductance L
i mH negligible
Medium temperature °C -40 … +80, -40 … +60 with oxygen
Ambient temperature °C -40 … +60 (T6)
CE-Conformity interference emission and immunity per EN 61326
Wiring L-connector (screw terminals up to 2.5 mm²)
Wiring protection protected against reverse polarity and overvoltage
Ingress protection IP 65 per EN 60529 / IEC 529
Wiring details,
2-wire
11265191.07 12/2012 GB/D/F/I
Earth, connected
to case 2)
U
B+/Sig+
Terminals 3, 4, 5 and 6: only for internal
application
0V/Sig-
WIKA operating instructions dierential pressure gauge models 712.15.160, 732.15.160
11
GB
8. Transmitter for level measurement
1) Only possible within 30 seconds of connecting the supply voltage
2) This connection must not be used
for equipotential bonding. The
instrument must be incorporated in
the equipotential bonding via the
process connection.