20 WIKA operating instructions pressure transmitter model O-10
GB
11613017.04 05/2013 GB/D/F/E
3. Specications
3.15 Electrical connections model O-10 (5)
Available connections
Electrical connection Ingress
protection
Wire cross-
section
Cable
diameter
Cable
lengths
Circular connector M12 x 1 (4-pin) IP 65 - - -
Metri-Pack series 150 IP 67
-
- -
Cable outlet, unshielded
1)
IP 67 0.14 mm
2
3.4 mm 2 m , 5 m
1) up to a maximum of 80 °C permitted
The stated ingress protection (per IEC 60529) only applies when plugged in using mating connectors
that have the appropriate ingress protection.
Mating connectors are not included in the delivery, but they are available as accessories.
Short-circuit resistance
S
+
vs. 0V
Reverse polarity protection
U
B
vs. 0V
Overvoltage protection
DC 36 V
Insulation voltage
DC 750 V
For special model numbers, e.g. O-10000, please note the specications stated on the delivery note.