Vista de página 7
WIKA operating instructions pressure gauge model PGS with reed contacts
GB
8
14000322.01 03/2011 GB/D/F/E
3. Specications
3. Specications
Limits for the contact rating with resistive load
Specications Model 851.3 Model 851.33
Used for instrument type PGS21.1x0, PGS23.1x0, PGS43.1x0,
PGS43HP.1x0, PGS63HP.1x0,
DPGS43.1x0, DPGS43HP.1x0,
APGS43.1x0
Contact design
1 change-over
contact
2 change-over
contacts
Type of contact bistable
Max. switching voltage AC/DC V 250
Min. switching voltage V not required
Max. switching current AC/DC A 1
Min. switching current mA not required
Max. carry current AC/DC A 2
cos ϕ 1
Contact rating W/ VA 60
Contact resistance (static) 100
Insulation resistance Ω 10
9
Breakdown voltage DC V 1000
Switching time incl.
contact chatter
ms 4.5
Contact material Rhodium
Switch hysteresis % 3 ... 5
The ingress protection to EN 60529 against external inuences depends
on the basic instrument and is found in the respective data sheet.
The limit switches can be used in the range from -20 to +70 °C. Where
the permissible temperatures for the basic instrument (see data sheet)
deviate from this range, the lower values are valid.
For further specications see the corresponding WIKA data sheet and
the order documentation.