20 WIKAoperatinginstructionstemperaturetransmitterT12
6.2.2 Sensors
Schematic representation / conguration
6. Commissioning, operation
Resistance thermometer (RTD) and resistance sensor
Itispossibletoconnectanresistancethermometer(e.g.to
DINEN60751)ina2-,3-or4-wireconnectionmethod.Congurethe
inputofthetransmittertomatchwiththeactualmethodofconnection
used.Otherwiseyouwillnotfullyexploitthepossibilitiesofconnec-
tionleadcompensationand,asaresult,possiblycauseadditional
measuringerrors(seechapter„6.2Conguration“).
Thermocouples (TC)
Makesurethatthethermocoupleisconnectedwiththecorrect
polarity.Iftheleadbetweenthethermocoupleandthetransmitter
needstobelengthened,onlyusethermalorcompensationcablein
accordancewiththetypeofthermocoupleconnected.
Conguretheinputofthetransmitterappropriatelyforthethermo-
coupletypeandthecoldjunctionactuallyused,otherwisemeasure-
menterrorsmaybecaused(seechapter"6.2Conguration").
Shouldthecoldjunctioncompensationbeoperatedwith
anexternalresistancethermometer(2-wireconnection),
connectthistoterminalsand.
Thermocouple
Resistancethermometer/
linearresistance
in
4-wire3-wire2-wire
CJCwith
external
Pt100/Ni100
1)
Input
mVsensor
4 ... 20 mA loop
CJC
internal
1)Connectsensor(Pt100/Ni100)forexternalcoldjunctioncompensationbetweenterminal1and4.