Vista de página 4
Installation with
spanner
WIKA Operating instructions pressure gauges Model 2 with 831 per ATEX
5
2064246 07/2009 GB/D/F
GB
3. Technical data and intended use
3. Technical data and intended use
Pressure limitation
Model 21X.20, 23X.50/30: Steady: full scale value
Fluctuating: 0.9 x full scale value
Short time: 1.3 x full scale value
Model 23X.36: Steady: full scale value
Fluctuating: 0.9 x full scale value
Short time: overload range
Process connection
In according with the general technical regulations for pressure gauges
(i.e. EN 837-2 „Selection and installation recommendations for pressure
gauges“).
When screwing gauges in, the force required for this must not be applied
through the case, but rather through the spanner flats (using a suitable tool)
provided for this purpose on the square shaft of standard connections.
Temperature effect
When the temperature of the measuring system deviates from the reference
temperature (+20 °C): max. ±0.4 %/10 K of full scale value
IP Ingress protection
Housing (EN 60 529/IEC 60 529)
Model 212.20: IP 54
Models 23X.X0/36: IP 65