11WIKA operating instructions, model CPH6600
14005103.03 11/2012 GB/D
GB
3. Specications
Available ranges and resolution
Pressure range and factor
Relative pressure -0.8 ... +2 bar -0.8 ... +10 bar -0.8 ... +20 bar
Overpressure safety 4 bar 13 bar 40 bar
Burst pressure 20 bar 20 bar 135 bar
Unit conversion factor
psi 1 30.000 150.00 300.00
bar 0.06894757 2.0684 10.3421 20.684
mbar 68.94757 2,068.4 10,342.1 20,684
kPa 6.894757 206.84 1,034.21 2,068.4
MPa 0.00689476 0.2068 1.03421 2.0684
kg/cm² 0.07030697 2.1092 10.5460 21.092
cmH
2
O (4 °C) 70.3089 2,109.3 10,546.3 21,093
cmH
2
O (20 °C) 70.4336 2,113.0 10,565.0 21,130
mmH
2
O (4 °C) 703.089 21,093 - -
mmH
2
O (20 °C) 704.336 21,130 - -
inH
2
O (4 °C) 27.68067 830.42 4,152.1 8,304.2
inH
2
O (20 °C) 27.72977 831.89 4,159.5 8,318.9
inH
2
O (60 °F) 27.70759 831.23 4,156.1 8,312.3
mmHg (0 °C) 51.71508 1,551.5 7,757.3 15,515
inHg (0 °C) 2.03602 61.081 305.40 610.81