GB
3211517.0707/2010GB/D
16 WIKAoperatinginstructionstemperaturetransmitterT12
Mounting example:
3173801.A
Mounting on the measuring insert
Mountthetransmitteronthecircularplateofthemeasuringinsert
usingtwocountersunkM3screwsperDINENISO2009.Appropriate
threadedinsertshavebeenpress-ttedintheundersideofthecase.
Assumingthecountersinkingiscarriedoutcorrectly,thepermissible
screwlengthcanbecalculatedasfollows:
l
max.
=s+4mm
with
l
max.
Lengthofscrewinmm
s Thicknessofcircular
plateinmm
3173801.B
max.4
Circularplate
l
s
6. Commissioning, operation
6. Commissioning, operation
Inhazardousareas,onlyusetemperaturetransmitters
thatareapprovedforthosehazardousareas.The
approvalismarkedontheproductlabel.
6.1 Mounting
6.1.1 Transmitter in head mounting version (model T12.10)
Thetransmittersforheadmountingaredesignedtobemounted
onameasuringinsertwithinaFormB,DINconnectionhead,with
extendedmountingspace.Theconnectionwiresofthemeasuring
insertmustbeapprox.50mmlongandinsulated.