Wika CPC6000 Manual do Utilizador

Consulte online ou descarregue Manual do Utilizador para Instrumentos de medida Wika CPC6000. WIKA CPC6000 User Manual Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir

Resumo do Conteúdo

Página 1 - CPC 6000

CPC 6000Operating InstructionsAutomated Pressure CalibratorAutomated Pressure Calibrator CPC 6000 PN 0017222001T

Página 2

Automated Pressure CalibratorCPC 6000 10 Mensor/WIKA Operating Instructions - CPC 6000NOTES

Página 3 - 1. General Information 7

Automated Pressure CalibratorCPC 6000 100 Mensor/WIKA Operating Instructions - CPC 600035 hectopascals hPa36megapascals MPa37 millimeters of water @ 2

Página 4 - 7. Remote Operation 45

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 10127 MSW 0°C 3.5% salinity 0.6838528 1.46230328OSI 16 0.062529 PS

Página 5 - 10. Calibration 89

Automated Pressure CalibratorCPC 6000 102 Mensor/WIKA Operating Instructions - CPC 600022 kPa 0.0001333220 7500.63625923 PA 0.1333220 7.50063625924 Dy

Página 6 - 11. Appendix 99

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 10318 MH2O 4°C 1.019744E-04 9.806378E+0319 mmHg 0°C 7.500636E-03 1

Página 7 - 1. General Information

Mensor Corporation201 Barnes DriveSan Marcos, Texas 78666Phone: 512.396.4200Fax: 512.396.1820Web site: www.mensor.comE-mail: [email protected]

Página 8

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 112. Safety Notices2.1 User ResponsibilitiesTo ensure safety, the

Página 9

Automated Pressure CalibratorCPC 6000 12 Mensor/WIKA Operating Instructions - CPC 60002.3 Warnings and Caution NoticesWARNING: HIGH PRESSURE! High pre

Página 10 - Automated Pressure Calibrator

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 133. Product Description3.1 General DescriptionThe Mensor CPC 6000

Página 11 - 2. Safety Notices

Automated Pressure CalibratorCPC 6000 14 Mensor/WIKA Operating Instructions - CPC 60003.2 FeaturesHere is a short list of significant features designe

Página 12

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 153.4 DisplayWhen the CPC 6000 is powered up it takes about one mi

Página 13 - 3. Product Description

Automated Pressure CalibratorCPC 6000 16 Mensor/WIKA Operating Instructions - CPC 6000Optional Display: The optional display is a window near the pres

Página 14

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 173.5 Electrical ModuleThe electrical module is illustrated below

Página 15

Automated Pressure CalibratorCPC 6000 18 Mensor/WIKA Operating Instructions - CPC 60003.6 Pneumatic ModulePneumatic modules come in two types and are

Página 16

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 193.7 Chassis AssemblyThe chassis assembly acts as the housing for

Página 17

Automated Pressure CalibratorCPC 6000 2 Mensor/WIKA Operating Instructions - CPC 6000This Warning symbol indicates that danger of injury for persons

Página 18

Automated Pressure CalibratorCPC 600020 Mensor/WIKA Operating Instructions - CPC 60003.8 Electrical Block DiagramFigure 3.8 - Electrical Block Diagram

Página 19

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 214. SpecificationsAccuracy specifications presented herein are ob

Página 20

Automated Pressure CalibratorCPC 600022 Mensor/WIKA Operating Instructions - CPC 6000Pressure Ranges Specific ranges within the range intervals are se

Página 21 - Bi-Directional

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 23Control Time Fast Mode: 10 seconds to stable flag for 10% FS ste

Página 22

Automated Pressure CalibratorCPC 600024 Mensor/WIKA Operating Instructions - CPC 60004.4 General SpecificationsSize 14.02” wide x 7.55” high x 12.42”

Página 23

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 255. Installation5.1 Unpacking the SystemIn addition to this manua

Página 24

Automated Pressure CalibratorCPC 600026 Mensor/WIKA Operating Instructions - CPC 60005.4 Rear PanelUp to eight pneumatic pressure ports are located ho

Página 25 - 5. Installation

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 275.5 Pressure ConnectionsWhen making up a connection to an o-ring

Página 26

Automated Pressure CalibratorCPC 600028 Mensor/WIKA Operating Instructions - CPC 60005.5.3 Exhaust PortIf sub-atmospheric control pressure is required

Página 27

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 296. Local Operation 6.1 GeneralThis section describes the procedu

Página 28

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 3TABLE OF CONTENTS

Página 29 - 6. Local Operation

Automated Pressure CalibratorCPC 600030 Mensor/WIKA Operating Instructions - CPC 6000Figure 6.3B is an example of a typical display after initializati

Página 30

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 31The current language selections available are:Language CountryEn

Página 31

Automated Pressure CalibratorCPC 600032 Mensor/WIKA Operating Instructions - CPC 6000Some pressure units can cause a number to be too long for the val

Página 32

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 33The current units are highlighted by a yellow background. Touch

Página 33

Automated Pressure CalibratorCPC 600034 Mensor/WIKA Operating Instructions - CPC 60006.3.8 Step Down [] and Step Up [] Keys As shown in Figure 6.3A,

Página 34

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 356.4 [Program] KeyThe [Program] key on the bottom left of the ope

Página 35

Automated Pressure CalibratorCPC 600036 Mensor/WIKA Operating Instructions - CPC 60006.4.1 Editing or Creating a ProgramTo edit or create a program, s

Página 36

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 37To change the name of the program, press the [Name] key. This di

Página 37

Automated Pressure CalibratorCPC 600038 Mensor/WIKA Operating Instructions - CPC 6000[Setpoint%] key: The [Setpoint%] key sets the control pressure at

Página 38

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 39The [>] key begins execution of the program at the first prog

Página 39

Automated Pressure CalibratorCPC 6000 4 Mensor/WIKA Operating Instructions - CPC 60006. Local Operation 296.1 General 296.2 Keys and Tabs 296.3 Disp

Página 40

Automated Pressure CalibratorCPC 600040 Mensor/WIKA Operating Instructions - CPC 6000The content of the three active displays are:[Pressure]: Shows no

Página 41

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 416.6.2 [Sensor] SetupTouch the [Sensor] tab and Figure 6.6.2A wil

Página 42

Automated Pressure CalibratorCPC 600042 Mensor/WIKA Operating Instructions - CPC 60006.7 Pump Regulator Control Limits: The control limits cannot be s

Página 43

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 43Rate: A rate (slewing speed) which best suits the user’s test re

Página 44

Automated Pressure CalibratorCPC 600044 Mensor/WIKA Operating Instructions - CPC 6000Rate: A rate (slewing speed) which best suits the user’s test req

Página 45 - 7. Remote Operation

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 457. Remote Operation7.1 Remote SetupUse the following screens to

Página 46

Automated Pressure CalibratorCPC 600046 Mensor/WIKA Operating Instructions - CPC 60007.1.2 [IEEE-488] Setup ScreenPress the [IEEE-488] tab to set up t

Página 47

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 477.1.4 [Ethernet] Setup ScreenPress the [Ethernet] tab to set up

Página 48

Automated Pressure CalibratorCPC 600048 Mensor/WIKA Operating Instructions - CPC 60007.3.1 Capability CodesSH1 Full source handshake capabilityAH1 Ful

Página 49

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 49Figure 7.4.1 - Serial Cable7.4.2 Command and Query FormatCommand

Página 50

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 57.4.7.3.1 Output Formats 657.4.7.3.2 Error Messages and Error Cod

Página 51

Automated Pressure CalibratorCPC 600050 Mensor/WIKA Operating Instructions - CPC 6000Always send commands in one of the following formats:1. [Command]

Página 52

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 517.4.5 Commands and QueriesTable 7.4.5 lists all of the current C

Página 53

Automated Pressure CalibratorCPC 600052 Mensor/WIKA Operating Instructions - CPC 600016 BR? <sp>{value}<cr><lf> Returns the B channe

Página 54

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 5339 Errorno? <sp>{string}<cr><lf> Returns CPC 6

Página 55

Automated Pressure CalibratorCPC 600054 Mensor/WIKA Operating Instructions - CPC 600059 Locale {locale code} Sets Languageand Country code.Locale Code

Página 56

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 5575 Peakmax? <sp>{value}<cr><lf> Returns the ma

Página 57

Automated Pressure CalibratorCPC 600056 Mensor/WIKA Operating Instructions - CPC 600094 Rsetpt? <sp>{value}<cr><lf> Returns the curr

Página 58

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 57118 StableWin {%fs value} Sets the stable window as a %FS.119 St

Página 59

Automated Pressure CalibratorCPC 600058 Mensor/WIKA Operating Instructions - CPC 60007.4.6 CPC 6000 Error CodesTable 7.4.6 - CPC 6000 Error CodesCodeS

Página 60

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 59E32 60h Invalid transducer selection INVALID TRANSDUCER SELECTIO

Página 61

Automated Pressure CalibratorCPC 60006 Mensor/WIKA Operating Instructions - CPC 600010.3 Pressure Standards 8910.4 Media8910.5 Calibration Setup8910.6

Página 62

Automated Pressure CalibratorCPC 6000 60 Mensor/WIKA Operating Instructions - CPC 60007.4.7 Remote EmulationThe Mensor PCS 400 is an earlier generatio

Página 63

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 6118 _pcs4 func stby <unitno> Instrument placed in standby m

Página 64

Automated Pressure CalibratorCPC 6000 62 Mensor/WIKA Operating Instructions - CPC 6000F$nnnnnnn6X RETURN NULL METER READING NOT SUPPORTEDF$nnnnnnn7X R

Página 65

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 637.4.7.3 SCPI CommandsSTATus : OPERation :CONDition? Returns a

Página 66

Automated Pressure CalibratorCPC 6000 64 Mensor/WIKA Operating Instructions - CPC 6000 :MODE OFF/GAS/LIQUIT Sets the head correction mode. :MEDIUM

Página 67

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 65 :AUTOvent ON/OFF/1/0 Sets autovent function on or off (automat

Página 68

Automated Pressure CalibratorCPC 6000 66 Mensor/WIKA Operating Instructions - CPC 6000-313 Calibration data not found-315 Configuration data not found

Página 69

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 67usage “I1” dummy function accepts the string and returns what wa

Página 70

Automated Pressure CalibratorCPC 6000 68 Mensor/WIKA Operating Instructions - CPC 6000case ‘U’: /*Units*/usage “UX” X = the unit number (ex. U16 = psi

Página 71 - 8. Options

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 6911 mmH2O12 cmH2O13 mH2O 20c14torr15Atm16 psi17 1b/ft2 not suppor

Página 72

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 71. General Information1.1 WarrantyAll products manufactured by Me

Página 73

Automated Pressure CalibratorCPC 600070 Mensor/WIKA Operating Instructions - CPC 6000NOTES

Página 74

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 718. OptionsThis section lists options available for the CPC 6000.

Página 75

Automated Pressure CalibratorCPC 600072 Mensor/WIKA Operating Instructions - CPC 60008.3 Calibration Sled KitThe calibration sled allows customers to

Página 76

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 738.5.1 Gauge Pressure Emulation KeyOn absolute units set the CPC

Página 77 - 9. Maintenance

Automated Pressure CalibratorCPC 600074 Mensor/WIKA Operating Instructions - CPC 60008.6.2 Screen SourceThe CPC 6000 has two frames in which channel i

Página 78

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 758.6.6 Screen Source A DeltaA is displayed on the left and is the

Página 79

Automated Pressure CalibratorCPC 600076 Mensor/WIKA Operating Instructions - CPC 60008.6.8 Delta Control LimitsThe default control limits for the Delt

Página 80

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 779. MaintenanceThe CPC 6000 was designed for maintenance-free ope

Página 81

Automated Pressure CalibratorCPC 600078 Mensor/WIKA Operating Instructions - CPC 60009.2 Troubleshooting the Pneumatic System of the CPC 6000PROCEED W

Página 82

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 79Figure 9.2.1C - Pneumatic Schematic, Pump Regulator

Página 83

Automated Pressure CalibratorCPC 60008 Mensor/WIKA Operating Instructions - CPC 6000These directives are designed to provide reasonable protection aga

Página 84

Automated Pressure CalibratorCPC 6000 80 Mensor/WIKA Operating Instructions - CPC 60009.2.2 Solenoid Valve Regulator The Solenoid Valve Regulator test

Página 85

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 81Figure 9.2.2B - Pneumatic Schematic, Solenoid Valve Regulator

Página 86

Automated Pressure CalibratorCPC 6000 82 Mensor/WIKA Operating Instructions - CPC 60009.2.2.1 Tuning a Solenoid Valve Regulator ModuleFigure 9.2.2.1 -

Página 87

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 839.2.2.1.2 Tuning Procedure1. Plug Measure/Control port.2. Conn

Página 88

Automated Pressure CalibratorCPC 6000 84 Mensor/WIKA Operating Instructions - CPC 60009.3.1 Transducer Removal1. VENT THE SYSTEM! Then turn off the p

Página 89 - 10. Calibration

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 859.3.2 Transducer InstallationTo replace a transducer first make

Página 90

Automated Pressure CalibratorCPC 6000 86 Mensor/WIKA Operating Instructions - CPC 60009.4 Spare Parts ListTable 9.4 lists the spare parts for the CPC

Página 91

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 879.5 Chassis InteriorPneumatic ModulesElectrical ModulePower Supp

Página 92

Automated Pressure CalibratorCPC 6000 88 Mensor/WIKA Operating Instructions - CPC 6000Figure 9.5B - Chassis Interior, Front View

Página 93

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 8910. Calibration10.1 GeneralThe CPC 6000 automatically adjusts th

Página 94

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 91.6.2 Calibration ServicesIn addition to servicing our own produc

Página 95

Automated Pressure CalibratorCPC 6000 90 Mensor/WIKA Operating Instructions - CPC 6000Figure 10.5 - Calibration Setup

Página 96

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 91To get to the Calibration screen press [Setup], [Sensor], and th

Página 97

Automated Pressure CalibratorCPC 6000 92 Mensor/WIKA Operating Instructions - CPC 60003. Enter the current password. As each number is pressed an ‘*’

Página 98

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 937. When satisfied that the new password is correct, and a writt

Página 99 - 11. Appendix

Automated Pressure CalibratorCPC 6000 94 Mensor/WIKA Operating Instructions - CPC 60003. Change the system password;4. Autozero all turndowns at one

Página 100

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 95For a gauge CPC 6000, this automatic function will:1. Vent the

Página 101

Automated Pressure CalibratorCPC 6000 96 Mensor/WIKA Operating Instructions - CPC 600010.8.4 2-Point CalibrationA two-point calibration adjusts both

Página 102

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 9710.9 Head Pressure CorrectionThe correction is accessed from the

Página 103

Automated Pressure CalibratorCPC 6000 98 Mensor/WIKA Operating Instructions - CPC 6000Local Gravity: Enter the local gravity acceleration value. If un

Página 104

Automated Pressure CalibratorCPC 6000 Mensor/WIKA Operating Instructions - CPC 6000 9911. Appendix11.1 Measurement UnitsThe Units command selects the

Comentários a estes Manuais

Sem comentários