7.4.6 SETUP additional menu items: Interface
7. Commissioning, operation
This menu allows the conguration of the interface. It can be congured for USB or
RS-232.
For the RS-232 interface there is a choice of 3 dierent baud rates with xed 8 used
data bits, no parity and 1 stop bit.
The USB connection can be switched on and o. In order to extend the battery life, the
USB connection should be deactivated when not in use.
7.4.7 SETUP additional menu items: CLEAR CalData
If the SETUP menu item "CLEAR CalData" is selected with the cursor and then the
SELECT button pressed 2x, then all stored calibration data will be deleted and reset.
The status of the deletion process is shown at the left of the lower Info
screen.
Select serial interface
Conrm selection
Change between interface
tape and baud rate