Wika S-10 Manual do Utilizador Página 6

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 35
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 5
1604457.14 GB/D/F/E 12/2009
10
WIKA Operating instructions/Betriebsanleitung/Mode d'emploi/Instrucciones de servicio S-10, S-11
1604457.14 GB/D/F/E 12/2009
11
WIKA Operating instructions/Betriebsanleitung/Mode d'emploi/Instrucciones de servicio S-10, S-11
Specifications Model S-10, S-11
Materials
Wetted parts (Other materials see WIKA diaphragm seal program)
Pressure connection» 316 Ti, S-11: O-ring: NBR
3)
{FPM/FKM}
Pressure sensor» 316 Ti (as of 40 bar / 600 psi 13-8 PH)
Case Stainless steel
Internal transmission fluid
4)
Synthetic oil
3)
O-ring made of FPM/FKM for model S-11 with integrated cooling element.
4)
Not for model S-10 with pressure ranges > 25 bar / 300 psi
Specifications Model S-10, S-11
Pressure ranges bar 0.1 0.16 0.25 0.4 0.6 1 1.6 2.5 4 6 10 16
Over pressure safety bar 1 1.5 2 2 4 5 10 10 17 35 35 80
Burst pressure bar 2 2 2.4 2.4 4.8 6 12 12 20.5 42 42 96
Pressure ranges bar 25 40 60 100 160 250 400 600 1000
1)
Over pressure safety bar 50 80 120 200 320 500 800 1200 1500
Burst pressure bar 96 400 550 800 1000 1200 1700
2)
2400
2)
3000
Pressure ranges psi 50 INWC 100 INWC 5 10 15 25 30 60 100
Over pressure safety psi 14.5 29 29 58 72 145 145 240 500
Burst pressure psi 29 35 35 69 87 170 170 290 600
Pressure ranges psi 200 300 500 600 1000 1500 2000
Over pressure safety psi 1160 1160 1160 1160 1740 2900 4600
Burst pressure psi 1390 1390 5800 5800 7970 11,600 14,500
Pressure ranges psi 3000 5000 8000 10,000
1)
15,000
1)
Over pressure safety psi 7200 11,600 17,400 17,400 21,750
Burst pressure psi 17,400 24,650
2)
34,800
2)
34,800 43,500
{Vacuum, gauge pressure, compound range, absolute pressure are available}.
1)
Only model S-10.
2)
For model S-11: The value specified in the table applies only when sealing is
realised with the sealing ring underneath the hex. Otherwise max. 1,500 bar /
21,000 psi applies.
(D) Mounting hole
1. Loosen the screw (1).
2. Loosen the cable gland (2).
3. Pull the angle housing (5), with the
terminal block (6) inside, away from the
instrument.
4. Using the head of a small screwdriver in
the mounting hole (D), lever the terminal
block (6) out of the angle housing (5).
In order not to damage the sealing of the angle housing, do not try to push the terminal
block (6) out using the screw hole (1) or the cable gland (2).
5. Ensure that the conductor outer diameter you select is matched to the angle housing’s
cable gland. Slide the cable through the cable gland nut (2), washer (3), gland seal (4) and
angle housing (5).
6. Connect the flying leads to the screw terminals on the terminal block (6) in accordance
with the pin-assignment drawing.
7. Press the terminal block (6) back into the angle housing (5).
8. Tighten the cable gland (2) around the cable. Make sure that the sealing isn’t damaged
and that the cable gland and seals are assembled correctly in order to ensure ingress
protection.
9. Place the flat, square gasket over the connection pins on the top of the instrument
housing.
10. Slide the terminal block (6) onto the connection pins.
11. Secure the angle housing (5) and terminal block (6) to the instrument with the screw (1).
Assembly of L-connector DIN EN 175301-803 Form A
(6)
(5)
(1)
(2)
(3)
(4)
Clamping nut,
Male connector,
Case with
pressure connection
Sealing
Female
connector
7. Starting, operation
GB
7. Starting, operation
GB
Vista de página 5