10 WIKA operating instructions, models CTH63I0 and CTH65I0
14120692.01 04/2015 EN/DE
2.7 Ex marking
DANGER!
Danger to life due to loss of explosion protection
Non-observance of these instructions and their contents may result in the
loss of explosion protection.
Observe the safety instructions in this chapter and further explosion
instructions in these operating instructions.
Observe the information given in the applicable type examination certifi-
cate and the relevant country-specific regulations for installation and
use in hazardous areas (e.g. IEC 60079-14, NEC, CEC).
Check whether the classification is suitable for the application. Observe the relevant
national regulations.
ATEX directive: 94/9/EC
Certificate no. TÜV 14 ATEX 143825 X
Ignition protection type: II 2 G Ex ib IIB T4 Gb
CE marking: 0158
Applied standards: EN 60079-0:2012
EN 60079-11:2012
EMC directive: 2004/108/EC
Applied standards: EN 61326 (group 1, class B)
II 2 G, Ex ib IIB T4 Gb
Marking Designation Meaning
CE marking Conformity markings in accordance with appendix
X of the 94/9/EC directive. The manufacturer shall
apply it before placing on the market.
II Instrument group The instrument may be used in hazardous areas,
except within the mining industry.
2G Instrument category With instrument category 2G, an potentially
explosive atmosphere involving gases (G) may
occur occasionally. The instrument can be used in
zone 1 and zone 2.
2. Safety