WIKA CTH65I0 Instruções de Operação Página 21

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 76
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 20
21
EN
21WIKA operating instructions, models CTH63I0 and CTH65I0
14120692.01 04/2015 EN/DE
5.4.2 Activating/deactivating display option differential temperature [Lin2]
(only for 2-channel instruments)
Press the ENTER/MENU key and select Lin2 using the arrow keys ▲▼. Then press
ENTER/MENU again. Now activate or deactivate the display “Differential temperature”
T
1
-T
2
using the arrow keys ▲▼.
If T
1
-T
2
is visible on the LC display, the differential temperature is active. Press ENTER/
MENU to confirm the desired setting.
Use ESC to return to the measuring mode.
To display differential temperature, both channels must be activated.
5.4.3 Calibration mode CAL
This measuring instrument allows to perform a simple calibration when replacing the
probes in order to compensate the manufacturer-specific tolerances of the probes and
guarantee a consistently high accuracy of the measuring chain.
The instrument has 3 different calibration modes:
CoFF: Standard characteristic curve
for example, for Pt100 resistance measurements DIN IEC 60751
oP1: Calibration by code
The 2 x 4-digit code is displayed on the handles of the probes, corresponds
to a 2-point calibration
oP2: Physical calibration
Calibration with reference standards: 1-point, 2-point or 3-point calibration
possible
The calibration coefficients are given in the calibration certificate.
CAL = calibrate
Press the ENTER/MENU key and select CAL using the ▲▼ arrow keys, then press
ENTER/MENU again. A small 1 appears on the left side of the display which indicates
MENU
Unit Lin2 CAL Chnl Lo6
5. Commissioning, operation