WIKA operating instructions models 53, 54, 55 11
3. Position the indicator as required, tighten the hexagon bolts and
slotted screws, and nally tighten the lock nut or union nut rmly.
When using thermowells, they must be lled with a thermal contact
medium in order to reduce the heat transfer resistance between the
outer wall of the sensor and the inner wall of the thermowell. The
working temperature of the thermal compound is -40 ... +200 °C.
WARNING!
Do not ll hot thermowells. There is a risk of the oil
spraying out!
Thermowell
Stem
required safety
clearance
loosening
Make sure to loosen the screws
on the opposite side as well!
CAUTION!
When using thermowells, please
ensure that the stem does not
touch the bottom of the thermowell
since, due to the dierent
expansion coecients of the
materials, the stem may become
buckled at the bottom of the
thermowell.
(Formula for the calculation of
the insertion length l
1
see the
respective thermowell's data sheet)
6. Commissioning, operation