Wika CPC6000 Manual do Utilizador Página 45

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 104
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 44
Automated Pressure Calibrator
CPC 6000
Mensor/WIKA Operating Instructions - CPC 6000 45
7. Remote Operation
7.1 Remote Setup
Use the following screens to set the operating parameters for the Ethernet, RS-232 serial port, and
IEEE-488.
Press the [Remote] key located on the bottom left corner of the screen, and a new display appears
with another set of tabs across the top as shown in Figure 7.1.1.
7.1.1 [Instrument] Setup Screen
Press the [Instrument] tab to set up available emulation modes. The default command set is Mensor.
Figure 7.1.1A - [Instrument] Setup Screen
[A] [B] (channel selection): This channel selection sets the active remote channel to A or B and is
useful for customers using an CPC 6000 to replace two single channel controllers. The user can simply
select the channel here first and then begin their normal program.
[Monitor] key: Press the [Monitor] key to bring up the remote monitor window which displays current
remote activity and syntax errors. This is helpful when troubleshooting programs.
Figure 7.1.1B - Monitor Window
Vista de página 44
1 2 ... 40 41 42 43 44 45 46 47 48 49 50 ... 103 104

Comentários a estes Manuais

Sem comentários