28 WIKA operating instructions, model CPH62I0
11221801.02 02/2011 GB/D/F/E
GB
6. Commissioning, operation
Viewing Recorded Measurements
Within the “LoGG-Stor“ function the measurements can be viewed directly on the
instrument display as well as through a computer (“Func-CYCL“ only allows for the data
to be viewed on computer). Press the ´SET/MENU´ key for 2 seconds: The rst menu
displayed is now “rEAd-LoGG“ (read logger data). By then pressing the ´TARA´ key, the
last measurement recorded will be displayed; pressing the ´TARA´ key again scrolls
between the dierent values related to this data point. Pressing the ´MIN´ key or ´MAX´
key selects dierent data points.
6.4.2 “Func-CYCL“: Automatic datalogging with adjustable cycle-time
The Logger cycle-time is adjustable (see Conguration). For example, setting “CYCL“ =
60: A measurement is made every 60 seconds.
When the slow measurement “rAtE-Slo“ is chosen, additionally a low power function is
available: “Lo.Po“. If “Lo.Po“ is on, the device only will take a measurement at the point
of time of the recording. In between the recordings the measuring is not active. This
decreases the power consumption enormously and therefore is recommended for long
time recordings where no mains adapter is available.
Max. number of measurements: CPH62I0-S1: 10,000
CPH62I0-S2: 4,000
(at max. 64 recording sequences)
Cycle-time: 1 ... 3,600 seconds (=1 h),
selectable from the conguration menu
A measurement contains:
Slow measuring rate (rAtE-SLo):
- Sensor 1: current value at that data point
- Sensor 1: Min-Peak, Max-Peak since last data point
- Sensor 2*: instantaneous value at that data point
- Sensor 2*: Min-Peak, Max-Peak since last data point
- Instantaneous dierence (Sensor 1 - Sensor 2)*: at that data point
- Min-Peak, Max-Peak of dierence (Sensor 1 - Sensor 2)*: Min-Peak, Max-Peak
since last data point
Fast measuring rate (rAtE-FASt, -P.dEt):
- Sensor 1: average since last data point
- Sensor 1: Min-Peak, Max-Peak since last data point
- Sensor 2*: average since last data point
- Sensor 2*: Min-Peak, Max-Peak since last data point
- Average dierence (Sensor 1 - Sensor 2)*: since last data point
- Min-Peak, Max-Peak of dierence (Sensor 1 - Sensor 2)*: Min-Peak, Max-Peak
since last data point
* Only for 2-channel version CPH62I0-S2