Wika CPH65I0 Manual do Utilizador Página 40

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 100
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 39
40 WIKA operating instructions, model CPH65I0
14061681.02 03/2013 GB/D
GB
6. Commissioning, operation
6.11.2 Calibrating a pressure-to-current transmitter
WARNING!
Observe the connection parameters prior to making a connection to the
instrument.
To calibrate a pressure-to-current transmitter (P/I), perform the following:
1. Connect the calibrator and the pump to the transmitter.
2. Apply pressure with the pump.
3. Measure the current output of the transmitter.
4. Ensure that the read value is correct. If it isn't, the transmitter must be adjusted.
Figure 5: CPH65I0 with a pressure transmitter
Vista de página 39
1 2 ... 35 36 37 38 39 40 41 42 43 44 45 ... 99 100

Comentários a estes Manuais

Sem comentários