8 WIKA operating instructions, model CPH65I0
14061681.02 03/2013 GB/D
GB
2. Safety
Connection values
Max. voltage U
o
= DC 7.14 V
Max. current I
o
= 1.12 mA
Max. power P
o
= 2 mW
Max. eective internal capacitance C
o
= 240 µF
Max. eective internal inductance L
o
= 1 H
Power supply circuit
Max. voltage U
i
= DC 30 V
Max. current I
i
= 80 mA
Max. power P
i
= 750 mW
Max. eective internal capacitance C
i
= 0 nF
Max. eective internal inductance L
i
= 0 mH
WARNING!
Further hazardous area safety instructions:
Observe the operating information and the relevant state regulations
concerning use in hazardous areas (e.g. EN IEC 60079-14).
The model CPH65I0 intrinsically-safe hand-held pressure calibrator has
been designed for use in hazardous areas. In these areas there is the
possibility of inammable or explosive gases being present.
The model CPH65I0 intrinsically-safe hand-held pressure calibrator has
been designed to be intrinsically safe. This means that connection with
other equipment that is in an intrinsically-safe circuit will not lead to arcs
capable of causing ignition, so long as the parameters are complied with.
Information/Approval for hazardous locations
WARNING!
Hazardous Areas
A hazardous area, as used in these operating instructions, refers to
an area made hazardous by the potential presence of ammable or
explosive vapours. These areas are also referred to as hazardous
locations.