18 WIKA Operating Instructions CTD 9300 Rev. 1
Temperature Dry Well Calibrators CTD 9300
2. Device Description
GB
When thermometers are calibrated that cannot be read out
electrically, a dialog box for entering the measured value will be
opened. In this case, an integrated instrument is not required.
Temperature Cycle Function
The temperature cycle function permits the recording of a
temperature curve that automatically switches between two
adjustable set values. In this case, the dwell time of the lower and
upper temperatures as well as the heating and cooling rates can be
set in K/min. The number of temperature changes is preset in the
beginning.
Series of Probes Testing Function
If multiple probes are to be calibrated at the same temperature, e.g.
when testing goods received, the "Check Series of Measurement"
mode must be selected. After the measured values have been
recorded, they can be output as a table and graphically in the form
of a histogram.
Fig. 6: Temperature ramp function