GB
14096788.01 07/2014 GB/D
12 WIKA operating instructions models TR31-3, TR31-K
ThermometerwithdirectsensoroutputwithPt100
(modelTR31-x-x-Px)andPt1000(modelTR31-x-x-Sx)
Temperature range without neck tube -50 ... +150 °C
(-58 ... +302 °F), with neck tube
-50 ... +250 °C (-58 ... +482 °F)
7)
Temperature at the connector or
at the directly connected cable
max. 85 °C (185 °F)
Measuring element
Pt100
(measuring current: 0.1 ... 1.0 mA)
Pt1000
(measuring current: 0.1 ... 0.3 mA)
Connection method
2-wire
3-wire
4-wire
Tolerance value of the measuring
element per IEC 60751
Class A
8)
Class B at 2-wire
Electrical connection
M12 x 1, 4-pin circular connector
Directly connected cable
7) Version with mineral-insulated sheathed cable can be used up to 300 °C (572 °F).
8) Class accuracy A only valid in the temperature range -30 ... +150 °C (-22 ... +302 °F) or
-30 ...+250 °C (-22 ... +482 °F), otherwise class B
For detailed specications for Pt sensors, see Technical information IN 00.17 at
www.wika.com.
Case
Material Stainless steel
Ingress protection
Case with connected connector
9)
or directly connected cable
Coupler connector, not connected
IP 67 and IP 69K per IEC 60529/
EN 60529
IP 67 per IEC 60529/EN 60529
Weight in kg ca. 0.2 ... 0.7
(depending on version)
Dimensions see “Dimensions in mm”
9) The stated ingress protection only applies when plugged in using mating connectors that have
the appropriate ingress protection.
3.Specications