Wika D-11-7 Manual do Utilizador Página 18

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 112
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 17
18 WIKA operating instructions pressure transmitter model D-1x-7
2478159.05 12/2011 GB/D/F/E
GB
6. Commissioning, operation
Overview of services
Output data
(from master)
Input data Explanation
Byte 1 Byte 2 Byte 3 Service request
(byte 6)
Service response 1
(byte 7)
Service response 2
(byte 8)
0x01 XX XX 0x01 ISE-SW AWE-SW Software version
0x02 XX XX 0x02 ISE-HW AWE-HW Hardware version
0x03 XX XX 0x03 MBA (value) Measuring range factor Initial pressure
0x04 XX XX 0x04 MBE (value) Measuring range factor End of measuring range
0x08 XX XX 0x08 Month Year Date of manufacture
0x10 0x10 Res.
0x12 0x12 Res.
0x14 0x14 Res.
0x15 0x15 Res.
0x1C 0x1C Res.
0x1D 0x1D Res.
0x21 XX XX 0x21 HI LO Read zero adjustment
0x22 XX XX 0x22 HI LO Read span correction
0x23 HI LO 0x23 HI LO Write zero adjustment
0x24 HI LO 0x24 HI LO Write span correction
0x42 0x42 Res.
0x44 0x44 Res.
0x54 XX XX 0x54 Sign Temperature value Read temperature
0x60 0x60 Res.
0x61 0x61 Res.
0x88 XX XX The 8 byte input data all display 0xFF Software reset
Vista de página 17