5 Set up
5.1 Adjustment system
1.1
2
3
1
Fig. 4: Indicating and adjustment elements
1 LC display
2 Indication of the menu item number
3 Adjustment keys
l [OK] key:
- Move to the me nu overview
- Conrm selected menu
- Edit parameter
- Save value
l [->] key to select:
- menu change
- list entry
- Select editing position
l [+] key:
- Change value of the parameter
l [ESC] key:
- interrupt input
- jump to the next higher menu
The sensor is adjusted via the four keys of the indicating and
adjustment module. The LC display indicates the individual menu
items. The functions of the individual keys are shown in the above
illustration. Approx. 10 minutes after the last pressing of a key, an
automatic reset to measured value indication is triggered. Any values
not conrmed with [OK] will not be saved.