Wika Pascal ET Manual do Utilizador Página 57

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 88
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 56
6 Commissioning, operation
WIKA Operating Instruction, Model Pascal ET and Pascal ET/IS 57
GB
The instrument shows:
Press “Zero Trim” or “Span Trim”, and the display shows (in this case Span Trim is selected):
The HART instrument is put in fixed mode generation (4 or 20 mA) and the REF channel is
measuring the true analog mA generation. To re-align to the correct 20 mA (or 4 mA) generation
press “Calibrate” button.
If needed follow the same procedure for “Zero Trim”.
Vista de página 56
1 2 ... 52 53 54 55 56 57 58 59 60 61 62 ... 87 88

Comentários a estes Manuais

Sem comentários