18 WIKA operating instructions, model CEP3000
14036068.02 08/2012 GB/D
GB
6. Commissioning, operation
6.2.3 Measuring temperature with a thermocouple
6.2.4 Measuring resistance and temperature with a resistance thermometer
The model CEP3000 oers three possibilities for measuring the resistance (Ω) and the
temperature with a resistance thermometer: 2-, 3- and 4-wire measurement. The highest
accuracy is obtained with a 4-wire connection.
Figures 4 - 6 show the connection of the model CEP3000 for the three dierent
measurement types.
1. In order to select the RTD mode, press the RTD key. Then, select the desired resist-
ance thermometer type with the RTD key.
2. The model CEP3000 must be operating in the input mode.
In the upper left corner of the LCD display, 2W, 3W or 4W are
displayed.
Select the desired connection conguration (2-wire, 3-wire or 4-wire) by using the up
and down arrow key.