Mensor/WIKA Operating Instructions Series CPG 2500 47
Digital Pressure Gauge
SERIES CPG 2500
ADDRESS address sp
<value>lf
Sets the device IEEE-488 address just as the ad-
dress key does in local operation. The address
can be any number from 0 through 30. Single digit
numbers can be preceded by a zero (01, etc.), but
the zero is not required. The command is address
sp <value><lf> where <value> is a number from 0
to 30.
Note: Allow at least 0.1 second between an ad-
dress change command and the next command to
allow the system to complete the change.
C not emulated
DEFAULT defaultlf Returns the CPG 2500 default values set at the
factory.
DIGITS digits sp <value>lf Sets the bus and display output resolution to ei-
ther five digits (full scale converted between 5,000
to 50,000 parts) or six digits (full scale between
50,000 to 500,000 parts). Send digits sp <value>lf
where <value> is 5 for five digits, or 6 for six digits
of resolution.
DIGITS? digits?lf The digits query returns the number of digits in the
output resolution.
Returns: <value><cr><lf> where <value> is either
5 or 6, same as above.
DISPLAY display sp<value>lf This command sets the display format.
Returns value<cr><lf> where <value> is:
0 normal
1 rate/second
2 rate/minute
3 hourly change
4 peak (min.)
5 peak (max.
6 rate/3 hour
7 tare (see below):
Display 7 is a toggle switch. If DISPLAY 7 is
sent to turn on the tare feature, DISPLAY 7
must be sent to disable it.