2059064.04 GB/D/F/E 06/2008
8
WIKA Operating instructions/Betriebsanleitung/Mode d'emploi/Instrucciones de servicio D-2X-9
2059064.04 GB/D/F/E 06/2008
9
WIKA Operating instructions/Betriebsanleitung/Mode d'emploi/Instrucciones de servicio D-2X-9
Specifications Model D-20-9, D-21-9
Pressure ranges bar 0.25 0.4 0.6 1 1.6 2.5 4 6 10 16
Over pressure safety bar 2 2 4 5 10 10 17 35 35 80
Burst pressure bar 2.4 2.4 4.8 6 12 12 20.5 42 42 96
Pressure ranges bar 25 40 60 100 160 250 400 600 1000
1)
Over pressure safety bar 50 80 120 200 320 500 800 1200 1500
Burst pressure bar 96 400 550 800 1000 1200 1700
2)
2400
2)
3000
{Vacuum, gauge pressure, compound range, absolute pressure are available}
1)
Only Model D-20-9.
2)
For model D-21-9: the value specified in the table applies only when sealing is
realised with the sealing ring underneath the hex. Otherwise max. 1500 bar applies.
Materials
Wetted parts (Other materials see WIKA diaphragm seal program)
Model D-20-9
*)
» Stainless steel
Model D-21-9» Stainless steel {Hastelloy}
O-ring: NBR {FPM/FKM or EPDM}
Case Stainless steel
Internal transmission fluid
3)
Synthetic oil {Halocarbon oil for oxygen applications}
{Listed by FDA for Food and Beverage}
3)
Not for D-20-9 with pressure ranges > 25 bar
Power supply UB UB in VDC 10 < UB 30
Signal output
CANopen protocol acc. to CiA DS-301 V. 4.02,
Device profile DS-404 V. 1.2
Power input W < 0.5 {with galvanic isolation < 0.7}
Communication services
LSS (CiA DSP 305, Version 1.1.1) Services
Configuration of device address and baud rate
Sync/Async, Node/Lifeguarding, Heartbeat
Filter
Individually programmable filter, to eliminate e. g. resonance frequen-
cies
Connect the instrument to earth via the pressure connection.
Operate the pressure transmitter with a shielded cable and earth the shield at least on
one side of the cable.
Use power supplies which guarantee reliable electrical isolation of the operating
voltage as per IEC/DIN EN 60204-1. Consider also the general requirements for PELV
circuits in accordance with IEC/DIN EN 60204-1.
Ingress protection per IEC 60529 (The ingress protection classes specified only apply
while the pressure transmitter is connected with female connectors that provide the
corresponding ingress protection).
Electrical connection
7. Starting, operation
GB
PIN assignment of connections according to CiA-DR 303-1
With integrated Y-piece
Circular connector
M 12x1, 5 pin, male
Circular connector
M 12x1, 5 pin, female
Circular connector
M 12x1, 5 pin, male
IP 67
Order code: M5
1 – Screen
2 – UB+ (CAN V+)
3 – UB- (CAN GND)
4 – Bus-Signal CAN-High
5 – Bus-Signal CAN-Low
1 – Screen
2 – UB+ (CAN V+)
3 – UB- (CAN GND)
4 – Bus-Signal CAN-High
5 – Bus-Signal CAN-Low
1 – Screen
2 – UB+ (CAN V+)
3 – UB- (CAN GND)
4 – Bus-Signal CAN-High
5 – Bus-Signal CAN-Low
IP 67
Order code: 2M
7. Starting, operation