Wika DPT-10 Manual do Utilizador Página 11

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 92
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 10
The conguration of the measuring cells diers depending on the
measuring range:
1
2
3
4
5
1
p
2
p
Fig. 7: Metallic measuring cell 10 mbar and 30 mbar - p
1
and p
2
process
pressures
1 Measuring element
2 Silicone diaphragm
3 Separating diaphragm
4 Filling oil
5 Integrated overvoltage arrester
1
p
2
p
1
2
3
4
Fig. 8: Metallic measuring cell from 100 mbar - p
1
and p
2
process pressures
1 Measuring element
2 Overload diaphragm/Middle diaphragm
3 Filling oil
4 Separating diaphragm
Power is supplied via the H1 Fieldbus. A two-wire cable according to
Fieldbus specication serves as carrier of both power and digital data
for multiple sensors. This cable can be operated in two versions:
l via an H1 interface card in the control system and additional power
supply
l via a Linking device with HSE (High speed Ethernet) and
additional power supply according to IEC 61158-2
Power supply and bus
communication
Dierential pressure transmitter DPT10 - metal measuring diaphragm Foundation Fieldbus 11
3 Product description
37246-EN-100120