The sensor diameter should be approx. 1 mm smaller than the bore diameter of the
thermowell or the blind bore, respectively.
Gaps of more than 0.5 mm between thermowell and the sensor will have a negative eect on
the heat transfer, and they will result in unfavourable response behaviour of the thermometer.
Sensor lengths
Sensor Ø in mm
Standard sensor lengths A (l
1
) in mm
6 50 100 150
8 - 100 150
Special lengths are possible.
Process connection
A compression tting enables simple, on-site adjustment to the required insertion length.
Compression tting
Material: stainless steel
G ¼ B male thread (not with Ø 8 mm sensors) or G ½ B
Delivery also possible without process connection.
Cable
Silicone, shielded, application range -40 ... +200 °C with EMC cable gland
Cable length to customer specication.
The sensor sleeve and case are connected by the cable shielding. Earthing on both ends
can lead to potential losses and indication inaccuracy.