14
WIKA operating instructions calibration system, model BCS10
14051046.01 10/2012 GB/D/F/E
GB
6. Commissioning, operation
6. Commissioning, operation
Commissioning may only be carried out by trained and qualied personnel. The calibra-
tion system was comprehensively checked ex works for tightness.
WARNING!
Only connect or disconnect test and calibration installations once the
system has been depressurised!
6.1 Switching on and o
To switch on the calibration system,
press the Power key (1). Press the
Power key again to switch o the
calibration system.
After switching on the version of the
rmware is indicated on the display for
approx. 1 second.
6.2 Zero point adjustment
To set the displayed value as zero
point, simply press the ZERO key (3).
The display changes immediately to the
0 value.
6.3 MIN/MAX value
The digital pressure gauge saves the
minimum and maximum pressure.
If the MAX/MIN key (4) is pressed once,
the maximum pressure from the memory is displayed. If the MAX/MIN key is pressed
again, the minimum pressure is displayed from the memory.
After 2 seconds, the display switches back to the operating mode. To clear the MAX/MIN
memory, press the MAX/MIN key (4) for at least 2 seconds until the message "CLr" is
displayed.
The bargraph display at the bottom of the display indicates the applied pressure relative
to the total measuring range.
Due to a possible Tare setting, the displayed pressure value can dier
from the applied pressure.
1
5
2
4
3