WIKA operating instructions calibration system, model BCS10
14051046.01 10/2012 GB/D/F/E
GB
6. Commissioning, operation
Switch on the calibration system.
The measurement setup shall be now as in the gure.
Actuate the pump handles several times until the
desired pressure is slightly exceeded.
To achieve the maximum pumping capacity, the spring above the knurled
nut must be relaxed.
For test items with low measuring ranges, the stroke path can be reduced
by turning the knurled nut. This will reduce the pressure increase per
stroke and minimize the risk of the overpressure.
Set the desirede pressure via the ne-adjustment
valve (arrow).
Normally, the test is carried out when the pressure falls
because the gas density measuring instruments are
generally specied and adjusted in such a way.
To reduce the pressure, turn the ne-adjustment valve
counterclockwise.
Further reduction of the pressure is possible by means
of the pressure relief valve.
To increase the pressure, turn the ne-adjustment
valve clockwise.
Use the digital pressure gauge to read the set pressure.