WIKA operating instructions pressure transmitter, model HP-2
11379180.03 09/2013 GB/D/F/E
GB
6. Commissioning, operation
6. Commissioning, operation
CAUTION!
Prior to commissioning, the pressure transmitter must be subjected to a
visual inspection.
Only use the pressure transmitter if it is in perfect condition with respect to
safety.
6.1 Mechanical mounting
The sealing faces at the instrument have to be
undamaged and clean.
When screwing the instrument in, the force
required to do this must not be applied through
the case, but only through the spanner ats
provided for this purpose and using a suitable
tool.
For the installation point, the valid values for
tightening torque and maximum pressure
should be taken from the high-pressure pipe
supplier's documentation.
When screwing in, do not cross the threads.
For information on tapped holes and welding sockets, see Technical information
IN 00.14 at www.wika.com.
6.2 Electrical mounting
The instrument must be grounded via the process connection.
Use the pressure transmitter with shielded cable, and ground the shield on at least
one end of the lead, if the lines are longer than 30 m (3- or 4-wire) or leave the build-
ing.
Select a cable diameter that matches the cable gland of the plug. Make sure that
the cable gland of the mounted plug has a tight t and that the seals are present and
undamaged. Tighten the threaded connection and check that the seal is correctly
seated, in order to ensure the ingress protection.
For cable outlets, make sure that no moisture enters at the cable end.
Observe the specications in accordance with chapter 3 “Specications”.
Spanner ats
Open-ended spanner
(spanner width 27)