Wika T32.xS Manual do Utilizador Página 12

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 60
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 11
GB
12
11583631.02 11/2010 GB/D/F/E
WIKA safety manual temperature transmitter T32.xS
Application (see table page 11):
Example 1
Sensor type Pt100, congured measuring range = -50 … +100 °C, so congured
measuring span = 150 K.
This is not smaller than 84 K. Thus the Total Safety Accuracy is 2 % FS,
thus 2 % * 150 K = 3 K,
and/or 2 % * 16 mA = 320 μA in terms of the current output
Example 2
Sensor type Pt100, congured measuring range = 0 … 50 °C, so congured
measuring span = 50 K
This is smaller than 84 K, thus the total safety accuracy is 2 K,
thus 2 K / 50 K = 4 %, and 4 % * 16 mA = 640 μA in terms of the current output
2.7 Conguration changes
WARNING!
During the conguration change, the safety function is not active! Safe
operation is only admissible with activated write protection (password).
Carry out conguration changes within the permissible specications in accordance
with chapter "2.1 Intended use in safety applications".
With the supplied conguration tools, the write protection, and other items, for the
model T32.xS is settable:
WIKA T32 Conguration Software
AMS
SIMATIC PDM
DTM (from Version DTM Beta version V1.0.2, January 2003) in conjunction with
operating software to the FDT/DTM Standard, e.g. PACTware, FieldMate
HART
®
hand-held terminal FC475, FC375, MFC4150
WARNING!
The safety function must be checked following any conguration
procedure.
2. Safety
Vista de página 11