WIKA operating instructions models TR12, TC12 15
5. Transport, packaging and storage
5.1 Transport
Check the instrument for any damage that may have been caused by transport.
Obvious damage must be reported immediately.
5.2 Packaging
Do not remove packaging until just before mounting.
Keep the packaging as it will provide optimum protection during transport (e.g. change in
installation site, sending for repair).
5.3 Storage
Permissible conditions at the place of storage:
Storage temperature:
Instruments without built-in transmitter: -40 ... +85 °C
Instruments with built-in transmitter: see operating instructions of the corresp. transmitter
Humidity: 35 ... 85 % relative humidity (no condensation)
Avoidance of exposure to the following factors:
Direct sunlight or proximity to hot objects
Mechanical vibration, mechanical shock (putting it down hard)
Soot, vapour, dust and corrosive gases
Potentially explosive environments, ammable atmospheres
Store the instrument in its original packaging in a location that fulls the conditions listed
above. If the original packaging is not available, pack and store the instrument as described
below:
1. Wrap the instrument in an antistatic plastic lm.
2. Place the instrument along with shock-absorbent material in the packaging.
3. If stored for a prolonged period of time (more than 30 days), place a bag containing a
desiccant inside the packaging.
WARNING!
Before storing the instrument (following operation), remove any residual media.
This is of particular importance if the medium is hazardous to health, e.g.
caustic, toxic, carcinogenic, radioactive, etc.
6. Commissioning, operation
WARNING!
When the thermometer is mounted, the temperature must not drop below the
allowed operating temperature (environment, medium) or exceed it, even when
taking convection and heat radiation into account!