Adjustment elements 4 keys
Protection rating
- unassembled IP 20
- mounted into the sensor without cover IP 40
Materials
- Housing ABS
- Inspection window Polyester foil
Voltage supply
Operating voltage
- Non-Ex instrument 12 … 36 V DC
- EEx-ia instrument 12 … 30 V DC
- Exd instrument 18 … 36 V DC
Operating voltage with lighted indicating and adjustment module
- Non-Ex instrument 20 … 36 V DC
- EEx-ia instrument 20 … 30 V DC
- EEx-d-ia instrument 20 … 36 V DC
Permissible residual ripple
- < 100 Hz U
ss
< 1 V
- 100 Hz … 10 kHz U
ss
< 10 mV
Load see diagram
1000
750
500
250
12 18
1614 20 22 24 26 28 30 32 34 36
W
V
3
1
2
Fig. 11: Voltage diagram
1 Voltage limit EEx-ia instrument
2 Voltage limit non-Ex/Exd instrument
3 Operating voltage
42 Process pressure transmitter IPT-1* Vers. 3.0 - metalic sensor • 4 … 20 mA
9 Supplement