7. Operating elements of the calibrator/micro calibration bath
7.4.2 Switching on automatic control
For certain tasks, it is advantageous to switch o control (e.g.
to make settings on the calibrator or micro calibration bath).
In calibration mode, press button P for approx. 5 secs., this
opens the main menu.
The upper PV display will show OPEr.
In the lower SV display, the LED SET will blink.
Press button 5 or button 6 until OFF is displayed.
Conrm with button P.
On the upper PV display, the current reference temperature
and OFF are indicated alternately.
On the lower SV display, the currently-selected set tempera-
ture is displayed.
CAUTION!
The control is now switched o and the
reference temperature will fall steadily,
unless it is corrected.
7.4.3 Switching on automatic control
The control is switched o when the following display is
shown:
On the upper PV display, the current reference temperature
and OFF are indicated alternately.
On the lower SV display, the currently-selected set tempera-
ture is displayed.
The control is switched back on by pressing button P for
approx. 5 secs., this opens the main menu.
The upper PV display will show rEG.
In the lower SV display, the LED SET will blink.
Conrm switching the control on by pressing button P.
CAUTION!
Control is now activated. The calibrator or
micro calibration bath will default to calibra-
tion mode and the set temperature will be
reached.
Display in main menu
Control OFF menu
Display with setting Control OFF
Display rEG
Display with setting Control OFF
Vista de página 25