8 WIKA operating instructions, models CTD9100, CTB9100, CTM9100
11263911.02 05/2014 GB/D
GB
3. Specications
3. Specications
3.1 Dry-well temperature calibrator, CTD9100 series
CTD9100-COOL CTD9100-165
Temperature range -55 ... +200 °C -35 ... +165 °C
Accuracy 0.15 ... 0.3 K 0.15 ... 0.25 K
Stability ±0.05 K ±0.05 K
Display resolution 0.01 to 100 °C, then 0.1 0.01 to 100 °C, then 0.1
Gradients, axially
1)
< 0.04 K at 200 °C < 0.04 K up to 100 °C
0.06 K up to 165 °C
Heating time incl. stabilisation
2)
15 min from 20 to 200 °C 12 min from 20 to 165 °C
Cooling time incl. stabilisation
2)
10 min from +20 to -20 °C 7 min from +20 to -20 °C
Immersion depth 150 mm 150 mm
Insert dimensions Ø 28 x 150 mm Ø 28 x 150 mm or Ø 60 x 150 mm
Voltage supply
Power supply AC 100 ... 240 V, 50/60 Hz AC 100 ... 240 V, 50/60 Hz
Power consumption 555 VA 375 VA
Power cord for Europe, AC 230 V for Europe, AC 230 V
Communication
Interface RS-485 RS-485
Case
Dimensions 215 x 305 x 425 mm (W x H x D) 215 x 305 x 425 mm (W x H x D)
Weight 11 kg 11 kg
CTD9100-450 CTD9100-650
Temperature range t
amb.
... 450 °C t
amb.
... 650 °C
Accuracy 0.3 ... 0.5 K 0.3 ... 0.8 K
Stability ±0.05 K at 100 °C
±0.1 K at 450 °C
±0.05 K at 100 °C
±0.1 K at 600 °C
Display resolution 0.01 to 100 °C, then 0.1 0.01 to 100 °C, then 0.1
Gradients, axially
1)
0.05 K at 100 °C
0.2 K at 450 °C
< 0.2 K at 100 °C
0.5 K at 600 °C
Heating time incl. stabilisation
2)
14 min from 20 to 450 °C 28 min from 20 to 600 °C
Cooling time incl. stabilisation
2)
58 min from 450 to 100 °C 60 min from 600 to 100 °C
Immersion depth 150 mm 150 mm
Insert dimensions Ø 60 x 150 mm Ø 28 x 150 mm
Voltage supply
Power supply AC 230/240 V, 50/60 Hz AC 230/240 V, 50/60 Hz
3)
(AC 100 ... 240 V, 50/60 Hz)
4)
Power consumption 2,000 VA 1,000 VA
Power cord for Europe, AC 230 V for Europe, AC 230 V
Communication
Interface RS-485 RS-485
Case
Dimensions 150 x 270 x 400 mm (W x H x D) 150 x 270 x 400 mm (W x H x D)
Weight 7.5 kg 8 kg
1) The gradient is understood to be the temperature change in the test well over the rst 40 mm from the bottom of the sleeve.
2) The reference thermometer with which the measurements are performed, has a diameter of 6 mm.
3) Instrument design available with multi-voltage power supply
4) AC 115 V power supply must be specied on the order, otherwise an AC 230 V one will be delivered.