46 WIKA operating instructions, models CTD9100, CTB9100, CTM9100
11263911.02 05/2014 GB/D
GB
13. Heating and cooling times
Infra-red insert
Heating up: Time Cooling: Time
-20 °C to -15 °C 0:30 min 150 °C to 125 °C 2:26 min
-15 °C to -10 °C 0:30 min 125 °C to 100 °C 2:52 min
-10 °C to 0 °C 1:00 min 100 °C to 75 °C 3:36 min
0 °C to 25 °C 2:42 min 75 °C to 50 °C 4:27 min
25 °C to 50 °C 3:06 min 50 °C to 25 °C 6:03 min
50 °C to 75 °C 3:26 min 25 °C to 0 °C 8:59 min
75 °C to 100 °C 4:12 min 0 °C to -10 °C 5:33 min
100 °C to 125 °C 5:38 min -10 °C to -15 °C 3:31 min
125 °C to 150 °C 9:49 min -15 °C to -20 °C 4:32 min
13.11 Model CTM9100-150 as surface temperature calibrator
Measuring conditions
All times refer to a Pt100 reference sensor with Ø 3 mm,
L = 300 mm
The reference sensor is inserted to a depth of 111 mm
All times are transition times and do not account for the
required settling time
The measurements were carried out at a room tempera-
ture of approx. 23 °C
It is controlled by an external reference sensor (3 x 300 mm)
Without cap, without covering
13.10 Model CTM9100-150 as an infrared black body source
Measuring conditions
Surface insert
Heating up: Time Cooling: Time
-20 °C to -15 °C 0:46 min 150 °C to 125 °C 3:11 min
-15 °C to -10 °C 0:45 min 125 °C to 100 °C 3:17 min
-10 °C to 0 °C 1:15 min 100 °C to 75 °C 3:51 min
0 °C to 25 °C 2:57 min 75 °C to 50 °C 5:02 min
25 °C to 50 °C 3:16 min 50 °C to 25 °C 6:58 min
50 °C to 75 °C 3:37 min 25 °C to 0 °C 11:55 min
75 °C to 100 °C 4:46 min 0 °C to -10 °C 9:19 min
100 °C to 125 °C 6:18 min -10 °C to -15 °C 7:44 min
125 °C to 150 °C 10:45 min -15 °C to -20 °C 12:35 min
All times refer to a Pt100 reference sensor with Ø 3 mm,
L = 150 mm
The reference sensor is inserted to a depth of 51 mm, and
is located under the front surface
All times are transition times and do not account for the
required settling time
The measurements were carried out at a room tempera-
ture of approx. 23 °C
It is controlled by an external reference sensor (3 x 300 mm)
Without cap, without covering