9WIKA operating instructions, models CTD9100, CTB9100, CTM9100
11263911.02 05/2014 GB/D
GB
3. Specications
Approvals and certicates, CTD9100 series
CE conformity
EMC directive 2004/108/EC, EN 61326 emission (group 1, class B) and interference immunity
(industrial application)
Low voltage directive 2006/95/EC, EN 61010-1, safety requirements for electrical equipment for measure-
ment, control and laboratory use
Certicate
Calibration 3.1 calibration certicate per DIN EN 10204
Option: DKD/DAkkS calibration certicate
For further specications see WIKA data sheet CT 41.28 and the order documentation.
3.2 Micro calibration bath, CTB9100 series
Model CTB9100-165 Model CTB9100-225
Temperature range -35 ... +165 °C t
amb.
... 225 °C, optional t
amb.
... 255 °C
Accuracy ±0.2 K ±0.3 K
Stability ±0.05 K ±0.05 K
Display resolution 0.1 °C 0.1 °C
Heating time incl. stabilisation
2)
30 min from 20 to 160 °C 20 min from 20 to 225 °C
Cooling time incl. stabilisation
2)
30 min from +20 to -20 °C 30 min from 225 to 50 °C
Immersion depth 150 mm 150 mm
Volume Approx. 0.6 litres Approx. 0.6 litres
Tank dimensions Ø 60 x 165 mm Ø 60 x 165 mm
Voltage supply
Power supply AC 100 ... 240 V, 50/60 Hz AC 230/240 V, 50/60 Hz
(AC 100/115 V, 50/60 Hz)
4)
Power consumption 375 VA 1,000 VA
Power cord for Europe, AC 230 V for Europe, AC 230 V
Communication
Interface RS-485 RS-485
Case
Dimensions 215 x 305 x 425 mm (W x H x D) 150 x 270 x 400 mm (W x H x D)
Weight 12 kg 7.9 kg
2) The reference thermometer with which the measurements are performed, has a diameter of 6 mm.
4) AC 115 V power supply must be specied on the order, otherwise an AC 230 V one will be delivered.
Accessories Model CTB9100-165 Model CTB9100-225
DC 200.05 silicone oil: -40 ... +130 °C,
FP* = 133 °C
from -35 ... +130 °C, very well usable not recommended
DC 200.10 silicone oil: -35 ... +160 °C,
FP* = 163 °C
from -35 ... +160 °C, well usable not recommended
DC 200.20 silicone oil: 10 ... 220 °C,
FP* = 232 °C
not recommended from 40 ... 225 °C, well usable
DC 200.50 silicone oil: 25 ... 250 °C,
FP* = 280 °C
not recommended from 80 ... 255 °C, well usable
* FP = Flash point