6.7 Saving the parameter adjustment data
It is recommended noting the adjusted data, e.g. in this operating
instructions manual and archive them afterwards. They are hence
available for multiple use or service purposes.
If DPT10 is equipped with an indicating and adjustment module, the
most important data can be read out of the sensor into indicating and
adjustment module. The procedure is described in the operating
instructions manual "Indicating and adjustment module" in the menu
item "Copy sensor data". The data remain there permanently even if
the sensor power supply fails.
If it is necessary to exchange the sensor, the indicating and adjustment
module is inserted into the replacement instrument and the data are
written into the sensor under the menu item "Copy sensor data".
58 Dierential pressure transmitter DPT10 - metal measuring diaphragm Probus PA
6 Adjustment with the indicating and adjustment module
37245-EN-100120