Max. torque screws mounting strap 30 Nm
Max. torque screws socket external housing 5 Nm (3.688 lbf ft)
Weight approx. 4.2 … 4 .5 kg (9.26 … 9.92 lbs), depending on
process fitting
Output variable
Output signal digital output signal, format according to IEEE-754
Sensor address 126 (default setting)
Current value 10 mA, ±0.5 mA
Dynamic behaviour output
Run-up time ≤ 20 s
Fig. 51: Presentation of the dead time t
1
and the time constant t
2
Version, nominal measuring range Dead time t
1
Time constant t
2
Basic version 10 bar and 30 bar 100 ms 250 ms
Basic version 100 mbar 100 ms 180 ms
Basic version 500 mbar 100 ms 180 ms
Basic version, 3 bar 100 ms 180 ms
Basic version 16 bar and 40 bar 100 ms 180 ms
Isolating diaphragm version, all nominal measuring
ranges
Depending on the isolat-
ing diaphragm
Depending on the isolating
diaphragm
Response time with Profibus PA
- cyclically approx. 10 ms
- acyclically approx. 50 ms
Damping (63 % of the input variable) 0 … 999 s, adjustable
74 Differential pressure transmitter DPT10 - metal measuring diaphragm • Profibus PA
11 Supplement
37245-EN-100120